Summary

We presented a comprehensive study of geometrical artifacts in a poorly calibrated OPT imaging system. We summarized our results in the form of a catalog by combining existing results on the shifted center of rotation and our own contribution to the 3D tilt of the rotation axis and the inaccurate rotation angles. In doing so, we are able to explain various types of mechanical artifacts, its appearance, cause, and properties, as well as the associated correction methods. This catalog serves as a reference for OPT practitioners to gain insight into their experimental setup and help them better calibrate their hardware system.

To fill the vacancy of a versatile computational method to account for all types of mechanical artifacts, we proposed an automatic calibration algorithm. It is based on a refined 3D mathematical model of the OPT imaging geometry that is able to characterize mechanical errors in the system. Moreover, the algorithm adapts to the large-size measurement datasets of OPT by performing the calibration on coarse scale to overcome the computational bottleneck, while the final reconstruction of the 3D volume is achieved on fine scale. Our multiscale calibration scheme was first validated on synthesized bead data where we simulated shifted center of rotation and imprecise rotation angles and successfully removed resulting artifacts. We then applied our algorithm to experimental datasets of a fluorescent textile fiber containing tilt errors, and successfully identified them to obtain an artifact-free reconstruction.

In practice, regularization may be introduced in the final reconstruction step. In principle, our geometric calibration framework can be adapted to other tomographic modalities, for instance, transmission OPT, optical coherence refraction tomography [118] or optical diffraction tomography [119].